Description


Page : 1797

Sondrestrom Lidar Operation Log


Operator(s) : Christian Have Pedersen
Purpose : No useful data is recorded. Test run with the purpose of reducing noise from scattered light in LIDAR room.
Date(s) : 070205 - 070205
File(s) : L20070116_001 - L20070205_001
Comments : Verified that scattered laser light is leaking into the ARCLITE receiver. The
effect of shielding critical points in the ARCLITE system was measured. There is
a noise level on the sodium channel even after light shielding the system and this
noise must be due to other factors than the scattered light in the LIDAR room.
Logfile : I0502070.log

Setup:0


Page : 1798
[Variables]

[0]
front end Iris=3.0mm
Rayleigh System=Off
Rayleigh Field stop=0.5mm
Rayleigh path attenuation=0dB
Dye System=Off
Sodium Field stop=0.5mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 1799

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : Christian Have Pedersen
Date(s) : 070205 - 070205
File(s) : L20070116_001 - L20070205_001
FilesTime [UT]CommentsCntSOpe
L20070116_00117:04:30Start of experiment0cp
L20070205_00017:23:45Dark LIDAR room. Reference lvl for the next couple of tests.00cp
L20070205_00017:24:57Roof hatch closed and lasers off.00cp
17:29:350cp
17:40:2500cp
17:50:1300cp
17:58:3600cp
18:05:420cp
18:16:1500cp
L20070205_00018:36:52Found out that frontend was blocked. Doing measurements00cp
from the start. Lights on Ref. level first.
L20070205_00118:50:58Frontend light.0cp
L20070205_00118:59:49Fronted light shielded00cp
L20070205_00119:12:40Telescope buttom light. Frontend shielded0cp
L20070205_00119:21:07Buttom light and partly shielded. (Shield not finished)0cp
L20070205_00119:28:04Just testing various stuff0cp
L20070205_00119:31:43End of experiment0cp