Description


Page : 9

Sondrestrom Lidar Operation Log


Operator(s) : J.Jørgensen
Purpose : Routine Data Collection
Date(s) : 980301 - 980302
File(s) : L9803012 - L9803021
Comments :
Logfile : L9803010.log

Setup:0


Page : 10
[Variables]
Rayleigh Power=570 mJ

[0]
Rayleigh System=On
Dye System=Off
Daystar=Out
Polerizer=Out
HalfWave Plates=Out

Logsheet:0


Page : 11

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : J.Jørgensen
Date(s) : 980301 - 980302
File(s) : L9803012 - L9803021
FilesTime [UT]CommentsCntSOpe
L980301223:52:10Start of experiment0JJ
23:54:00Dye system keeps dying0JJ
23:57:00Something wrong with dye system! going to run with0JJ
rayl. 90 channel only.
L980301200:16:00While checking on other optical equip. the rayl. system 0JJ
had shut off. Started it again and everything seems to work
just fine.
L980301200:30:00HG adj.0JJ
L980301200:36:58HG and conv. mirror adj7770JJ
L980301201:05:07Most have set the dye SR430 wrong during startup. 0JJ
Noticed the rayl. was waiting after it finished its cycle.
Connected the 30 Hz trigger to the dye SR30 which
seemed to fix the problem.
L980301201:33:55HG adj0JJ
L980301202:23:488430JJ
L980301303:00:21Nice counts today8350JJ
L980301303:54:34Noticed SR430 had stopped with a frozen profile on0JJ
the display.
L980302003:56:43Running again new filename. I have no Idea how long the 0JJ
system had been frozen The program was still running
with yellow light blinking and averything.8010JJ
L980302004:58:06HG adjust7410JJ
L980302005:14:13HG adjust0JJ
L980302107:03:09End of experiment8350JJ