Description


Page : 18

Sondrestrom Lidar Operation Log


Operator(s) : E.Gudmundsson
Purpose : Routine data collection
Date(s) : 980321 - 980322
File(s) : L9803210 - L9803213
Comments : Had some problems starting up. The Rayleigh heat exchanger did not show
anything on led's or temperature display. After some time I found a blown 0.5A
fuse in the control circuit located under the control panel. Started taking data
with setup 0 at 22:39 UT and the Rayleigh system was on at 01:26 UT (setup 1)
Logfile : L9803210.log

Setup:0


Page : 19
[Variables]
Rayleigh Power=630mJ
Dye Power=80mJ

[0]
Rayleigh System=Off
Dye System=On
Daystar=Out
Polerizer=Out
HalfWave Plates=Out

[1]
Rayleigh System=On
Dye System=On
Daystar=Out
Polerizer=Out
HalfWave Plates=Out

Logsheet:0


Page : 20

Sondrestrom Lidar Operation Log


Log Sheet 1 of 2
Operators : E.Gudmundsson
Date(s) : 980321 - 980322
File(s) : L9803210 - L9803212
FilesTime [UT]CommentsCntSOpe
L980321022:39:05Start of experiment0EG
L980321022:39:08Ups! changed dye trigger to 10Hz and rec per scan to 6000EG
L980321022:44:54Dye system tuned to 589.158 0EG
L980321022:58:12Convergence mirror adjusted dye0EG
L980321023:04:32Convergence mirror adjusted dye0EG
L980321023:07:01Convergence mirror adjusted0EG
L980321023:16:03Dye system tuned to 589.1590EG
L980321023:18:30Convergence mirror adjusted0EG
L980321000:53:02System paused0EG
L980321000:57:50Resumed Dye system tuned to 589.158 0EG
L980321101:09:46Dye system tuned to 589.1590EG
L980321101:14:26Dye system tuned to589.1600EG
L980321101:20:47Dye system tuned to589.1610EG
L980321101:26:30Rayleigh systen up and running now781EG
L980321101:38:24Ups! Opened receiver, light got in3341EG
L980321101:44:07Rotated 90% PMT in its cooler (been removed since last time6221EG
L980321101:56:22Adjusted Rayleigh Convergence mirror6301EG
L980321102:00:37Rotated PMT6421EG
L980321102:10:32Convergence mirror adjusted Dye & Rayleigh HG adjusted5351EG
L980321102:26:46Tweeked Rayleigh HG5701EG
L980321103:18:37Dye system tuned to 589.1615251EG
L980321103:21:54Dye system tuned to 589.1625861EG
L980321203:22:56Counts5911EG
L980321203:27:38Tuned dye to 589.1635471EG

Logsheet:1


Page : 21

Sondrestrom Lidar Operation Log


Log Sheet 2 of 2
Operators : E.Gudmundsson
Date(s) : 980322 - 980322
File(s) : L9803212 - L9803213
FilesTime [UT]CommentsCntSOpe
L980321204:20:09Convergence mirror adjusted Rayleigh4831EG
L980321204:35:49Rayleigh convergence and HG tweeked3941EG
L980321205:35:46Convergence mirrors adjusted and Rayleigh HG5741EG
L980321205:39:23Tweeked dye HG output power 83mJ (did not pause)5341EG
L980321306:24:40End of experiment5621EG