| Files | Time [UT] | Comments | Cnt | S | Ope |
|---|
| L2010 | 21:25:13 | Start of experiment | | 0 | EG |
| L2010 | 21:26:38 | Nice and clear out there, Nice sodium counts | | 0 | EG |
| L2010 | 21:28:00 | !UPS forgot to change the recs per scan to 600 | | 0 | EG |
| L2010 | 21:28:21 | Done changing | | 0 | EG |
| L2010 | 21:43:18 | Tuned Dye HG energy was down to 50mJ ended at 85mJ after | | 0 | EG |
| | tweeking | | | |
| L2010 | 21:57:23 | sloping background. plaied wiht cables and convergence | | 0 | EG |
| 22:00:23 | done playing | | 0 | EG |
| L2011 | 22:13:23 | Incremented and Changed discriminator level to -4.0mV | | 0 | EG |
| L2011 | 22:17:21 | Changed disc level to -7mV | | 0 | EG |
| L2011 | 23:47:39 | Sodium Convergence mirror adjusted, sealed the dye laser | | 0 | EG |
| | chimney to try to block air currents. | | | |
| L2011 | 23:56:48 | Looking at the displays this sealing seems to have had an | | 0 | EG |
| | effect. | | | |
| L2011 | 00:19:31 | Sodium Convergence mirror adjusted | | 0 | EG |
| L2011 | 03:11:36 | After Eggert blocked the holes at the bottom of the chimney | | 0 | JJ |
| | the counts became a lot more stable | | | |
| L2011 | 03:14:32 | End of experiment | | 0 | JJ |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |