| Files | Time [UT] | Comments | Cnt | S | Ope |
|---|
| L3041 | 20:19:14 | Start of experiment | 5977 | 0 | EG |
| L3042 | 20:45:02 | Incremented filenumber, Sodium system running now | | 0 | EG |
| L3042 | 20:47:15 | Getting a good sodium profile SZA=90.9 | 5941 | 1 | EG |
| L3042 | 23:02:58 | Rayleigh HG and Conv. mirror adj | 2865 | 1 | JJ |
| L3042 | 23:49:01 | Rayleigh Convergence mirror adjusted | 2636 | 1 | KR |
| L3042 | 02:22:39 | Seeing one-integration spikes in Rayleigh counts by ~50% | 4067 | 1 | KR |
| | every so often. Not sure why. | | | |
| L3042 | 06:29:53 | Rayleigh Convergence mirror adjusted | 4974 | 1 | KR |
| L3042 | 06:54:59 | Incremented, opening up the iris to see if the signal goes up | 7131 | 1 | EG |
| L3050 | 06:57:02 | The iris is now at ~1.5mm | 6923 | 2 | EG |
| L3050 | 07:10:12 | Got signal gain on the Rayleigh, sodium not affected by this | 9824 | 2 | EG |
| | moved the sodium chopping up in altitude as it seemed | | | |
| | that we were getting signal induced noise in the system | | | |
| L3050 | 09:22:34 | Closing the iris down again | 7409 | 2 | EG |
| L3050 | 09:24:58 | iris has been closed, incremented | 7538 | 1 | EG |
| L3051 | 09:26:10 | counts | 2729 | 1 | EG |
| L3051 | 10:08:51 | Sodium Convergence mirror adjusted | 1018 | 1 | EG |
| L3051 | 10:12:09 | Rayleigh Convergence mirror adjusted | 1543 | 1 | EG |
| | It's prooving difficult to converge the Rayleigh with the min | | | |
| | opening on iris. Convergence is flaky and changes when you | | | |
| | close the door to the converging mirror!!!! | | | |
| L3051 | 10:13:30 | counts | 2193 | 1 | EG |
| L3051 | 10:47:52 | SZA=88.5 still seeing sodium on the raw profile | 4793 | 1 | EG |
| | Rayleigh S/N ~4000/700 | 4878 | | |