Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson, A.Merkel
Purpose : data set used for tweeking the reciever. Testing the reciever after the installation of the mirror mount
Date(s) : 000525 - 000525
File(s) : L5250 - L5252
Comments : We started data taking at 2:02:37 but we were checking the count rate with the
new
folding mirror in the reciever. The last mirror before the PMT in the 532
channel. In the time between the start of data and the first data file increment
several changes were done to try and characterize the count rate. The PMTs were
turned off, the half waveplate alignment of the transmitter was checked to be ok.
We knew that the mirror may have an effect on the count rate but we went
through other checks to make sure that it was just the mirror. The count rate is
down from 9000 to 2000 with the addition of the mirror. This is not a good data
set for science purposes
Logfile : l0005250.log

Setup:0


Page : 538
[Variables]
Dye Power=100 mJ
Rayleigh Power=540 mJ

[0]
Iris=min_opening
Rayleigh System=On
Dye System=On
Daystar=Out
Polarizer=In
HalfWave Plates=In
Rayleigh Filter=2:532(.5nm)

Logsheet:0


Page : 539

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : E.Guðmundsson, A.Merkel
Date(s) : 000525 - 000525
File(s) : L5250 - L5252
FilesTime [UT]CommentsCntSOpe
L525002:02:35Start of experiment0AM
L525103:14:46Increment file0AM
L525103:25:40Gained signal by tweeking HG25350AM
03:27:37converged Dye laser27830AM
L525103:39:42Paused data collection to change to the old mirror mount26820AM
04:09:27Contunued after changing mirror, using laser as reference0AM
L525204:54:31Calibrated mirror setup, now running new data0
04:58:35Still tweeking reciever. Not a good data set for science0
05:26:38Stoping taking data from Sodium channel0
05:32:48Taking Sodium data again0
05:46:52End of experiment0