Description


Page : 661

Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson
Purpose : Routine data collection
Date(s) : 001002 - 001002
File(s) : L10020 - L10021
Comments :
Logfile : l0010020.log

Setup:0


Page : 662
[Variables]
Dye Power=93mJ
Rayleigh Power=540mJ

[0]
Iris=min_opening
Rayleigh System=On
Dye System=On
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)

[1]
Iris=min_opening
Rayleigh System=On
Dye System=Off
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)

Logsheet:0


Page : 663

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : E.Guðmundsson
Date(s) : 001002 - 001002
File(s) : L10020 - L10021
FilesTime [UT]CommentsCntSOpe
L1002000:40:01Start of experiment00EG
L1002000:50:23Tuned sodium system to 589.158nm67440EG
L1002000:53:34Rayleigh Convergence mirror adjusted66500EG
L1002001:04:36Sodium window is building up frost on the inside!0EG
L1002001:21:02I put a 60w lamp underneeth the sodium chimey to see if81430EG
the frost goes away
L1002002:23:56Clouds in the path now18240EG
L1002002:31:41Working on the sodium window, no sodium data now18401EG
L1002003:03:32Put some insulation on the window holder.12261EG
L1002003:03:51Incremented4521EG
L1002103:05:02Counts5000EG
L1002103:26:05The clouds are here to stay, shutting down.0EG
L1002103:26:35End of experiment250EG