Description


Page : 972

Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson
Purpose : routine data collection
Date(s) : 020219 - 020219
File(s) : L20020219_000 - L20020219_003
Comments : light haze to begin with. thickening towards the end of the run. Not the bestof
conditions.
Logfile : l0202190.log

Setup:0


Page : 973
[Variables]
Dye Power=90mJ
Rayleigh Power=490mJ

[0]
Iris=0.6mm
Rayleigh System=On
Dye System=Off
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=In
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

[1]
Iris=0.6mm
Rayleigh System=On
Dye System=On
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=In
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 974

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : E.Guðmundsson
Date(s) : 020219 - 020219
File(s) : L20020219_000 - L20020219_003
FilesTime [UT]CommentsCntSOpe
L20020219_00001:04:21Start of experiment0EG
01:08:51running with rayleigh system only, waiting for na to warm.0EG
01:40:37Haze is attenuationg the beam0EG
L20020219_00101:45:23incremented to 0010EG
L20020219_00201:50:45Incremented to 002, and started the sodium channel.1EG
03:37:49Tuned from 589.161 to 589.1581EG
L20020219_00305:32:16End of experiment1EG