Description


Page : 1445

Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson
Purpose : Routine data collection and finalyzing some tweeking for summer
Date(s) : 040511 - 040512
File(s) : L20040511_000 - L20040512_001
Comments :
Logfile : l0405110.log

Setup:0


Page : 1446
[Variables]
Dye Power=45mJ
Rayleigh Power=420mJ

[0]
front end Iris=3.0mm
Rayleigh System=On
Rayleigh Field stop=2.5mm
Rayleigh path attenuation=0dB
Dye System=Off
Sodium Field stop=2.5mm
Etalon=Out
Daystar=8A_In
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

[1]
front end Iris=3.0mm
Rayleigh System=On
Rayleigh Field stop=2.5mm
Rayleigh path attenuation=0dB
Dye System=Off
Sodium Field stop=4.0mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

[2]
front end Iris=3.0mm

Setup:1


Page : 1447
Rayleigh System=On
Rayleigh Field stop=3.0mm
Rayleigh path attenuation=0dB
Dye System=On
Sodium Field stop=4.0mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 1448

Sondrestrom Lidar Operation Log


Log Sheet 1 of 2
Operators : E.Guðmundsson
Date(s) : 040511 - 040512
File(s) : L20040511_000 - L20040512_000
FilesTime [UT]CommentsCntSOpe
L20040511_00023:14:38Start of experiment0EG
23:17:40Starting by verifying RX halfwave position0EG
L20040511_00123:55:30Starting taking some data to get a hold of status, plates aligned0EG
00:12:14counts3000EG
L20040511_00201:29:05removing daystar from receiver path3001EG
L20040511_00201:32:05Taking data without the daystar1EG
L20040511_00202:10:45Finally found the sodium beam, starting sodium data collection16042EG
L20040511_00202:15:59Sodium SR430 froze, reset while Rayleigh was taking data15982EG
Looks like the power cycle may have corrupted a record
L20040511_00202:18:17I'd better increment the file2EG
L20040512_00002:20:04From the profile, sodium signal is looking good - considering low 15982EG
laser energy. Field stop seems to require some attention
Will not open the receiver now - I don't want to interupt
the Rayleigh collection yet by opening up the receiver
L20040512_00002:27:05Rayleigh counts looking good. I've been looking for the 15892EG
6min variability in the profile to profile history but I'm not
able to see it???
L20040512_00002:51:23looked like the counts on the Rayleigh are dropping so I2EG
Tried to converge, was not better.
L20040512_00003:34:40There seems to be a thin haze layer attenuating the signal 12522EG
a bit.
L20040512_00004:31:27closing down, going to take a look at the iris settings before11512EG
L20040512_00004:36:02Closed the Rayleigh field stop as far down as possible, tweeked13262EG
the XY position slightly.

Logsheet:1


Page : 1449

Sondrestrom Lidar Operation Log


Log Sheet 2 of 2
Operators : E.Guðmundsson
Date(s) : 040512 - 040512
File(s) : L20040512_000 - L20040512_001
FilesTime [UT]CommentsCntSOpe
L20040512_00004:42:23Sodium field stop opeing reduced a bit and reconverged13162EG
L20040512_00004:45:13Sodium channel reconverged again13202EG
L20040512_00004:49:36Could probably crank the Sodium field stop further down,13262EG
did not see a noticable drop in background as with the
Rayleigh field stop.
L20040512_00104:50:57taking data for a few minutes before shutting down for13262EG
tonight.
L20040512_00105:02:44End of experiment13282EG