Description


Page : 1681

Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson, M.McCready, Claus Johannsen
Purpose : Routine data collection
Date(s) : 050908 - 050909
File(s) : L20050908_000 - L20050909_006
Comments :
Logfile : l0509080.log

Setup:0


Page : 1682
[Variables]
Dye Power=77mJ
Rayleigh Power=550mJ

[0]
front end Iris=not_installed
Rayleigh System=On
Rayleigh Field stop=1.5mm
Rayleigh path attenuation=0dB
Dye System=On
Sodium Field stop=1.5mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 1683

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : E.Guðmundsson, M.McCready, Claus Johannsen
Date(s) : 050908 - 050909
File(s) : L20050908_000 - L20050909_006
FilesTime [UT]CommentsCntSOpe
L20050908_00023:00:32Start of experiment10300EG
L20050908_00023:22:45Sodium channel dead- took record and run off10470EG
turned record and run on again.--no good
Turned SR430 off, toggled run and record on the software
Sodium channel taking data again - it looks like the SR430
is freezing and not the software. Need to repeat this
on next occurence. Software was not reset but kept running
L20050908_00000:23:28Sporadic sodium10000EG
L20050908_00000:49:19sporadic sodium still lingering9690MM
L20050908_00000:52:44tuned sodium from 589.160 to 589.1589710MM
L20050909_00003:54:44had message: couldn't open \\umimmak\...\loginfo.txt file,9980MM
but system is still running and data seems okay.
L20050909_00406:39:14restart Na SR430 -- didn't help9420MM
06:43:26restarted it again, to no avail -- called Eggert0MM
L20050909_00507:01:11had to restart FormMain application9410MM
L20050909_00610:44:38End of experiment4150CJ