Sondrestrom Lidar Operation Log


Operator(s) : E.Guðmundsson
Purpose : Run coordinated with the IS Radar in meteor mode
Date(s) : 060103 - 060104
File(s) : L20060103_000 - L20060104_001
Comments : Changed dye and made some alignments on pumping of the sodium laser. Pump
beam seemed to have changed alignment a tad causing loss of pumping in preamp
and main amp dye cell. problem is fixed Dye laser energy now 95mJ with a good
beam profile.
Logfile : l0601030.log

Setup:0


Page : 1706
[Variables]
Rayleigh Power=490
Dye Power=95mJ

[0]
front end Iris=not_installed
Rayleigh System=Off
Rayleigh Field stop=1.5mm
Rayleigh path attenuation=0dB
Dye System=On
Sodium Field stop=1.5mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

[1]
front end Iris=not_installed
Rayleigh System=On
Rayleigh Field stop=1.5mm
Rayleigh path attenuation=0dB
Dye System=On
Sodium Field stop=1.5mm
Etalon=Out
Daystar=Out
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 1707

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : E.Guðmundsson
Date(s) : 060103 - 060104
File(s) : L20060103_000 - L20060104_001
FilesTime [UT]CommentsCntSOpe
L20060103_00020:50:00Start of experiment00EG
21:10:22focused on getting dye laser on line, had to change dye due 0EG
to low energy as well a do some minor pump beam alignments.
looking good now with dye enregy at 95mJ./pulse and a really
good looking profile.
L20060103_00121:11:32working on getting the Rayleigh system on line.0EG
21:12:09some sligh haze at ~7km0EG
L20060103_00121:26:06OOPS! bumped the sodium convegence mirror out of align.0EG
while starting up the Rayleigh system. reconverged
L20060103_00121:37:30Reyligh system on line, incremented0EG
L20060103_00121:37:46counts1EG
L20060103_00222:24:54Something not ok with the Rayleigh system.1EG
L20060103_00222:37:07Looks like the Reyleigh PMT has bought it!1EG
No signal as measured directly on the PMT. DC level high
Supply current arround the normal 500uA. HV ok.
Could be a short somewhere in the dynode chain
L20060103_00223:16:26restart data acq.01EG
L20060103_00323:59:53Skies getting worse1EG
L20060103_00300:00:07The Rayleigh system fault could be a bad amp?01EG
L20060103_00300:39:48Reyligh laser has been turned off-data acq on Rayleigh1EG
lidar stopped.
L20060103_00300:57:06low altitude cloud cover getting denser, still able to see good01EG
sodium profiles.
L20060104_00104:16:27End of experiment01EG