Sondrestrom Lidar Operation Log


Operator(s) : Katelynn, Greer, Katelynn Greer
Purpose : Determining whether there is a fluctuation in Laser Energy during polarization flipping
Date(s) : 080715 - 080715
File(s) : L20080712_000 - L20080712_000
Comments : This was an experiment to determine whether there was a fluctuation of power in
between the flipping of polarizations. Another problem that has been
encountered is a dust tube between the amplifier and SHG keeps coming out of
place. This is evidenced in lower power readings. Although no noticable
difference was seen during polarization flipping, the TX plate was not properly
resetting to zero. This likely need to be manually reset, and as the thumb screw
that I can see is not progressing the dial, I am awaiting further instructions from
Matt Hayman.
Logfile : l0807150.log

Setup:0


Page : 2018
[0]
front end Iris=not_installed
Rayleigh System=On
Rayleigh Field stop=0.5mm
Rayleigh path attenuation=0dB
Dye System=Off
Sodium Field stop=0.5mm
Etalon=In
Daystar=4A_In
Polarizer=In
HalfWave Plates=Tracking
Rayleigh Filter=2:532(.5nm)
Rayleigh seeding=off

Logsheet:0


Page : 2019

Sondrestrom Lidar Operation Log


Log Sheet 1 of 1
Operators : Katelynn Greer
Date(s) : 080715 - 080715
File(s) : L20080712_000 - L20080712_000
FilesTime [UT]CommentsCntSOpe
L20080712_00021:23:54Start of experiment0KRG
Initial SP Energy: 5050 mJ0
Powermeter in Beam Path
21:28:44Begin Flipping Routine: Constant Energy of 4100 mJ0KRG
21:36:30End Flipping Rountine0KRG
Just Sun Tracking: 4100 mJ
21:39:35Ending Data Acq program, 1/2 wave plates should = 00KRG
Transmit plate = 94
21:44:24Try starting and ending experiment again0KRG
TX plate not resetting to 0
Unable to find manual control
22:07:34End of experiment0KRG