| Files | Time [UT] | Comments | Cnt | S | Ope |
|---|
| L9806140 | 00:28:40 | Start of experiment | | 0 | JJ |
| L9806140 | 00:37:23 | HG adj. | 526 | 0 | JJ |
| L9806140 | 01:03:22 | HG adj. | 647 | 0 | JJ |
| L9806140 | 01:06:01 | Something is wrong, the background is going up when it | | 0 | JJ |
| | should be going down. | | | |
| L9806140 | 01:37:21 | Pausing to look at RX halfwave plate | | 0 | JJ |
| L9806140 | 01:40:07 | Running again. Wondering if the RX halfwave plate is | | 0 | JJ |
| | moveing in the wrong direction since they moved it to the | | | |
| | other side of the beamsplitter???? The is no easy way to | | | |
| | test it though without reprogramming. | | | |
| L9806140 | 01:42:12 | Ups forgot the blocking filter was in place | | 0 | JJ |
| L9806140 | 01:43:01 | Okay back to filter 3 | | 0 | JJ |
| L9806140 | 01:48:41 | Disconnected both TX and RX halfwave trackers to see if | | 0 | JJ |
| | the background will start going down | | | |
| L9806140 | 01:58:15 | Trackers connected again. Background started to go down | | 0 | JJ |
| | so something is wrong with the halfwave plate tracking. | | | |
| L9806140 | 01:59:14 | Going to adj. the halfwave plates | | 0 | JJ |
| L9806140 | 02:15:22 | Adjusted halfwave plates and was able to lower the back- | | 0 | JJ |
| | ground quiet a bit!! | | | |
| L9806140 | 02:19:25 | HG + conv. mirror adj. | 565 | 0 | JJ |
| L9806141 | 02:29:53 | That seemed to have helped. The background is decreasing | 652 | 0 | JJ |
| | now as it should. | | | |
| L9806141 | 03:57:56 | HG adj. | 721 | 0 | JJ |
| L9806142 | 05:33:03 | End of experiment | | 0 | JJ |