Description


Page : 95

Sondrestrom Lidar Operation Log


Operator(s) : J.Geddes
Purpose : Routine data collection and experiment comparing new 532nm filter and Daystar filter effectiveness.
Date(s) : 980717 - 980718
File(s) : L9807170 - L9807173
Comments : At the very beginning of the run, the Daystar filter was out, but the 532nm filter
was not in place. Therefore, the Rayleigh PMT saturated. The PMT voltage was
turned down, the Daystar filter was replaced, and the PMT began collecting data
once the voltage was turned up again.

During the first cycle of removing the Daystar filter and subsequently replacing it,
the chopper wheel was accidentally turned off. It was turned back on again
immediately. No data was compromised.

During the run, data collection was periodically stopped, the Daystar filter was
replaced with a 532nm filter, and several minutes of data in this configuration
were
obtained. In this way a measure of the relative performance of these components
at diffferent SZAs was found.

The skies were clear throughout the night.
Logfile : l9807170.log

Setup:0


Page : 96
[Variables]
Rayleigh Power=525mJ

[0]
90% PMT Iris=not_installed
Rayleigh System=On
Dye System=Off
Daystar=In
Polerizer=In
HalfWave Plates=Tracking
Filter wheel=3:Open

[1]
90% PMT Iris=not_installed
Rayleigh System=On
Dye System=Off
Daystar=Out
Polerizer=In
HalfWave Plates=Tracking
Filter wheel=2:532/0.5nm

[2]
90% PMT Iris=not_installed
Rayleigh System=On
Dye System=Off
Daystar=Out
Polerizer=In
HalfWave Plates=Tracking
Filter wheel=1:532/1nm

Logsheet:0


Page : 97

Sondrestrom Lidar Operation Log


Log Sheet 1 of 2
Operators : J.Geddes
Date(s) : 980717 - 980718
File(s) : L9807170 - L9807170
FilesTime [UT]CommentsCntSOpe
L980717022:49:03Start of experiment30J.G
L980717022:59:14System is working now. Disregard first 10 minutes of data. Daystar filter was missing.00J.G
L980717023:03:50HG adjusted4840J.G
L980717023:07:21HG adjusted5730J.G
L980717023:12:03HG adjusted5700J.G
L980717023:25:29System paused at end of record 35. Daystar filter was removed, and 5A 532nm filter was put in its place. 6631J.G
L980717023:28:23Data collection resumed. SZA is 81.3 degrees.1J.G
L980717023:32:27System paused at end of record 39. Daystar filter was replaced, and 5A 532nm filter was removed. 42490J.G
L980717023:35:16Data collection resumed. SZA is 81.8 degrees.0J.G
L980717023:40:49System paused at end of record 45. Daystar filter was removed, and the 5A 532nm filter was put in its place.. 1J.G
L980717023:43:35Data collection resumed. SZA is 82.5 degrees.1J.G
L980717023:46:22System paused at end of record 48. Daystar filter was replaced, and the 5A 532nm filter removed.0J.G
L980717023:48:38Data collection resumed. SZA is 82.9 degrees.0J.G
L980717023:54:29System paused at end of record 54. Daystar filter was removed and replaced with 5A 532nm filter.1J.G
L980717023:56:32Data collection resumed. SZA is 83.5 degrees.41171J.G
L980717023:59:28Data collection paused after record 57. The 5A 532nm filter was removed, and the Daystar was replaced.0J.G
L980717000:01:55Data collection resumed. SZA is 83.9 degrees.8130J.G
L980717000:08:42System paused at end of record 64. Daystar filter was removed, and 5A 532nm filter was put in its place.1J.G
L980717000:10:45Data collection resumed. SZA is 84.5 degrees.1J.G
L980717000:13:14System paused at end of record 67. 5A 532nm filter was removed, and the Daystar filter was replaced.0J.G
L980717000:15:39Data collection resumed. SZA is 84.8 degrees.0J.G
L980717000:24:36System paused after record 76. Daystar filter was removed, and 5A 532nm filter was put in its place.1J.G
L980717000:26:46Data collection resumed. SZA is 85.6 degrees.1J.G
L980717000:30:19System paused after record 80. Daystar filter was replaced, and the 5A 532nm filter was removed.0J.G

Logsheet:1


Page : 98

Sondrestrom Lidar Operation Log


Log Sheet 2 of 2
Operators : J.Geddes
Date(s) : 980718 - 980718
File(s) : L9807170 - L9807173
FilesTime [UT]CommentsCntSOpe
L980717000:32:56Data collection resumed. SZA is 86.0 degrees.0J.G
L980717000:45:43System paused after record 93. Daystar filter was removed and replaced with the 5A 532nm filter.1J.G
L980717000:47:48Data collection resumed. SZA is 86.9 degrees.1J.G
L980717000:52:15System paused after record 98. 5A 532nm filter was replaced with 10A 532nm filter.2J.G
L980717000:53:33Data collection resumed. SZA is 87.2 degrees.2J.G
L980717000:57:29System paused after record 102. 10A 532 nm filter was removed, and the Daystar filter was replaced.0J.G
L980717000:59:25Data collection resumed. SZA is 87.6 degrees.0J.G
L980717001:09:29System paused at end of record 111. Daystar was removed, and 5A 532nm filter was put in its place.1J.G
L980717001:11:42Data collection resumed. SZA is 88.2 degrees. 1J.G
L980717104:01:42Computer has recorded a "slow or missing triggers" warning.31591J.G
L980717307:45:19End of experiment1J.G