| Files | Time [UT] | Comments | Cnt | S | Ope |
|---|
| L9809210 | 23:40:50 | Start of experiment | 2838 | 0 | EG |
| L9809210 | 23:44:23 | Rayleigh HG adjusted | 2866 | 0 | EG |
| L9809210 | 00:10:51 | Sodium Convergence mirror adjusted | 2849 | 0 | EG |
| L9809210 | 00:11:13 | !! looks like a sporadic peak at 92.0 km | 2832 | 0 | EG |
| L9809210 | 00:59:40 | UPS SR430 frozen reebooted sr430 | 2846 | 0 | EG |
| L9809210 | 01:08:09 | Rayleigh Convergence mirror adjusted | 2211 | 0 | EG |
| L9809210 | 01:25:12 | Rayleigh Convergence mirror adjusted | 2175 | 0 | EG |
| | My Rayleigh counts have been juping around a bit | | | |
| L9809210 | 01:25:42 | Counts | 2406 | 0 | EG |
| L9809211 | 03:25:30 | Rayleigh HG adjusted | 2406 | 0 | EG |
| | Rayleigh Convergence mirror adjusted | | | |
| L9809211 | 03:30:20 | I think the Rayleigh system is behaving oddly (countwize) | 1580 | 0 | EG |
| L9809211 | 03:35:33 | Pausing to check power | 2068 | 0 | EG |
| L9809211 | 03:39:58 | Resuming, could not get the energy above 530mJ !! | 1880 | 0 | EG |
| L9809211 | 03:50:39 | I've been checking, Lamp counts are aorund 27E6 since last | | 0 | EG |
| L9809211 | 03:51:33 | change | 1547 | 0 | EG |
| L9809211 | 03:51:39 | I don't want to run with the system as it is, closing down | | 0 | EG |
| L9809211 | 03:52:13 | the rayleigh system | 1575 | 0 | EG |
| L9809220 | 03:53:52 | Resuming with setup 1 incremented | 1045 | 1 | EG |
| L9809220 | 03:55:29 | Wavemeter dead on at 589.158 since startup | | 1 | EG |
| L9809220 | 04:02:05 | !! noticed that after i turned the Rayleigh system off, the | | 1 | EG |
| L9809220 | 04:03:17 | sodium counts fell to virtually nothing at zero altitude | 0 | 1 | EG |
| | This is probably crosstalk from the Rayleigh channel but | | | |
| | does this effect the data? | | | |