| Files | Time [UT] | Comments | Cnt | S | Ope |
|---|
| L9810270 | 01:02:57 | Start of experiment | 2502 | 0 | EG |
| L9810270 | 01:04:48 | I'm having some problems with the dye system | | 0 | EG |
| L9810270 | 01:19:59 | Pausing data collectionto swap HV units | | 0 | EG |
| L9810270 | 01:21:33 | Resuming , no difference | 2278 | 0 | EG |
| L9810270 | 01:28:30 | Removed the gate from the sodium PMT, signal goes out | 2269 | 0 | EG |
| L9810270 | 01:31:11 | Put the Gate on again, adjusted convergence | 2317 | 0 | EG |
| L9810270 | 01:34:16 | Pausing and stopping the chopper | 2288 | 0 | EG |
| L9810270 | 01:34:17 | Resuming with chopper off running on line freq. | 2335 | 0 | EG |
| | No differance in the signal | | | |
| L9810270 | 01:36:15 | Pausing again to set the system on chopper again | | 0 | EG |
| L9810270 | 01:37:37 | Resuming | 2480 | 0 | EG |
| L9810270 | 01:50:17 | I'm baffled ??? The PMT currend is ok. everthing looks in | | 0 | EG |
| | order. Can it be the PMT?? | | | |
| L9810270 | 02:09:10 | Rayleigh Convergence mirror adjusted | 2172 | 0 | EG |
| L9810270 | 02:13:12 | opened for the PMT filter wheel in open position, ahha! | 2130 | 0 | EG |
| L9810270 | 02:17:24 | incremented filenumber | | 0 | EG |
| L9810271 | 02:17:40 | The wavelength was way off around half a nanometer! | 2176 | 0 | EG |
| | During initial tuning the wavemeter had jumped down ca 0.5 | | | |
| | nm but then up again. When I left it the first time it was | | | |
| | dead on (or so I thought) but when checking now again the | | | |
| | display indicated 588.670nm! Tunied again and the same | | | |
| | wavelength jumps were present during tuning?? | | | |
| | The wavemeter is stable now at 589.158nm now | | | |
| L9810271 | 02:57:06 | This was sudden! Clouds all over the place! closing down | 2 | 0 | EG |